| Version 16 (modified by , 17 years ago) ( diff ) |
|---|
Ion Source
We use a Peabody Scientific Duoplasmatron ion source to produce our H- beam.
Wien Filter
Right after the extraction gap of the source is a Wien filter, or ExB filter, or charge/mass (q/m) filter, where crossed electrical and magnetic fields let only ions with a selectable ratio q/m pass.
Source Setup
Source Maintenance
Filament Coating
Inside the source is a platinum mesh electron gun which starts a plasma. The mesh has a special coating which wears out
after about 2-3 months of running (depending on amount of use). As time goes on and the coating wears out the source
will need more arc current and source gas (H2) to maintain beam magnitude. When the coating is really worn out you will not
be able to start a plasma.
Water Filter Change
Another paramater to keep on eye on is the current drawn by the primary electrode. If the power source of the primary electrode
(labeled Beam Energy) is over 1.2 mA, then this could be a sign that the cooling water has too many impurites and has a
decreased electrical resistance. The water must be changed and possibly the de-magnetizing filter.
The filter can be obtained by contacting Peabody Scientific (Frank Schmurra: 978.535.0444).
The filter is a bag of beads in a sealed cloth bag. DO NOT OPEN THE BAG!!! The bag must be
intact when it is put into the system. Go to the blue-grey box under the bender and open it.
Be careful, there is electric equipment everyhwere, shut off all close by electic equipment, the turbo and rough pump
are ok to leave on.Inside is the cooling system for the source. Using a pan, drain the water out of the system.
There is a clear plastic cylindrical filter holder, unscrew the outside cap carefully so to not spill the water in it
on top of the motor that pumps the water through the cooling lines. The old filter bag is sitting in this piece in a
mesh basket. Simply take out the old one, and place the new filter inside the mesh. The old one can be thrown out with the
regular trash. Make sure that the rubber gasket is in place and screw the filter holder back on.
Re-connect the water pipes and add distilled water in the PVC pipe top , it should only take 0.5 to 0.75 gallons (2-3 Liters) at most.
There is a flow meter interlock in this cooling system, the source power will shut off if it does not detect fluid flowing.
When turning the cooling system on, hold down the power button for a minute or so, this makes sure all the air pockets get pushed
out of the system which could have shut down the power.
[[Image(https://docs.astro.columbia.edu/attachment/wiki/SavinGroup/MergedBeams/SourceSection/091006%20second%20leg%20pics%2C%20coolin%20system%20inside.jpg?format=raw
, 50%)]]
Attachments (4)
-
Ion Source.jpg
(105.4 KB
) - added by 17 years ago.
Ion Source
- 091006 second leg pics.jpg (24.3 KB ) - added by 17 years ago.
-
091006 second leg pics, coolin system outside.jpg
(61.1 KB
) - added by 17 years ago.
Source Cooling Box
-
091006 second leg pics, coolin system inside.jpg
(50.8 KB
) - added by 17 years ago.
Source Cooling System Box Inside
Download all attachments as: .zip

