    3535There is a clear plastic cylindrical filter holder, unscrew the outside cap carefully so to not spill the water in it
    3636on top of the motor that pumps the water through the cooling lines. The old filter bag is sitting in this piece in a
    37 mesh basket. Simply take out the old one, and place the new filter inside. The old one can be thrown out with the
     37mesh basket. Simply take out the old one, and place the new filter inside the mesh. The old one can be thrown out with the
    3838regular trash. Make sure that the rubber gasket is in place and screw the filter holder back on.
    39 Re-connect the water pipes and add distilled water in the PVC pipe top , it should only take one gallon (3.8 Liters)? at most.
     39Re-connect the water pipes and add distilled water in the PVC pipe top , it should only take 0.5 to 0.75 gallons (2-3 Liters) at most.
    4040There is a flow meter interlock in this cooling system, the source power will shut off if it does not detect fluid flowing.
    4141When turning the cooling system on, hold down the power button for a minute or so, this makes sure all the air pockets get pushed