Unmodified
Added
Removed
Modified
  • SavinGroup/MergedBeams/FirstLeg

    v6 v7  
    33= First Leg of the Beamline =
    44
    5 last modified: Apr. 25 2007
     5last modified: Sept. 20 2007
    66
    77
    88[[BR]]
    9 Here we take the beam from the [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/IonSource Ion source] and [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Einzel focus] it, [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Steerer steer] it, [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Slits shape] it, determine its [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#BPM profile], and then either [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Bender deflect] it or guide it into a [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Cup Faraday cup]. Here's some details on the single elements:
     9Here we take the beam from the [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/IonSource Ion source] and [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Einzel focus] it, [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Steerer steer] it, [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Slits shape] it, determine its [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#BPM profile], and then either [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Bender deflect] it or guide it into a [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/FirstLeg#Cup Faraday cup]. Here's the current setup in the Lab (right to left: source, first Einzel lens, [http://en.wikipedia.org/wiki/Mass_spectrometry#Mass_analyzer ExB filter], second Einzel lens, steerers, alignment bellows, beam defining slits, beam profile monitor, retractable Faraday-cup, chamber for spherical deflector, another beam profile monitor, and a static Faraday-cup.)
     10[[Image(first leg.jpg, 50%)]]
     11
     12Here are some details on the single elements:
    1013
    1114== Miniature Einzel Lens == #Einzel
    12 Our Einzel lens sits right after the exit of the [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/IonSource#ExB Wien Filter] that cleans up our beam from whatever other ions/electrons are extracted from the plasma in the source. To catch the beam at a small diameter, we need to put the Einzel Lens very close to the Wien Filter exit aperture. Following advice from Michael Rappaport, we have used ceramic balls as spacers between the three elements. A 45 mil (1.15mm) diameter wire is [http://en.wikipedia.org/wiki/Tig_welding TIG welded] to the top of the central lens element. Next time we should use a smaller diameter wire: the welder likes thicker wires because they don't just evaporate when he does the weld, we like a thicker diameter because it gives the wire stability, but we will still have to bend it. In the DIN100 Tee we used we had only little access to bend the wire at the right position, and it was a bit of a fight to make the connection to the electrical feedthrough nice and sound. The lens has now been conditioned to 25kV, which should be more than enough for everything we plan to do with it. Note that the only flange that can be opened without damaging anything is the downstream-flange of the tee (because of the electrical connection)! Here's some pics:
     15Our Einzel lens sits right after the exit of the [https://docs.astro.columbia.edu/wiki/SavinGroup/MergedBeams/IonSource#ExB Wien Filter] that cleans up our beam from whatever other ions/electrons are extracted from the plasma in the source. To catch the beam at a small diameter, we need to put the Einzel Lens very close to the Wien Filter exit aperture. Following advice from Michael Rappaport, we have used ceramic balls as spacers between the three elements. A 45 mil (1.15mm) diameter wire is [http://en.wikipedia.org/wiki/Tig_welding TIG welded] to the top of the central lens element. Next time we should use a smaller diameter wire: the welder likes thicker wires because they don't just evaporate when he does the weld, we like a thicker diameter because it gives the wire stability, but we will still have to bend it. In the DIN100 Tee we used we had only little access to bend the wire at the right position, and it was a bit of a fight to make the connection to the electrical feedthrough nice and sound. The lens has now been conditioned to 25kV, which should be more than enough for everything we plan to do with it. Note that the only flange that can be opened without damaging anything is the downstream-flange of the tee (because of the electrical connection)! Here are some pics:
    1316
    1417
    … …  
    3336
    3437== XY Deflector == #Steerer
    35 The steerers a scaled down version of a design we were given by Aaron Covington (many thanks again!). They fit into one 6" tee; according to our simulations they will deflect our beam nicely with a few 100V defletion voltages.
     38The steerers a scaled down version of a design we were given by Aaron Covington (many thanks again!). They fit into one 6" tee; according to our simulations they will deflect our beam nicely with a few 100V defletion voltages. We have two pairs of horizontal and vertical deflection electrodes, allowing accurate placing of the beam rather than just introducing an angle.
    3639
    3740Right after mounting:
    … …  
    5053
    5154== 90° Deflector == #Bender
    52 We deflect the beam by 90° with a spherical deflector such that no UV photons from the ion source can reach the interaction region of the experiment. We decided to use a spherical deflector because (in a perfect world) it doesn't screw up our beam shape and needs only low voltages to operate as compared to other options like cylindrical deflectors or quadrupole deflectors. Again, the design was originally inspired by a design from Aaron Covington that he gave us, but he used spherical plates that we found would be too hard to obtain/machine. So we changed the design considerably (will post pictures once we have them). There are two holes in the outer deflection electrode; one allows us to see the beam after the deflector if we apply no voltages (i.e. the beam travels straight through) and helps with the alignment of the assembly, the other is only for alignment purposes.
     55We deflect the beam by 90° with a spherical deflector such that no UV photons from the ion source can reach the interaction region of the experiment. We decided to use a spherical deflector because (in a perfect world) it doesn't screw up our beam shape and needs only low voltages to operate as compared to other options like cylindrical deflectors or quadrupole deflectors. Again, the design was originally inspired by a design from Aaron Covington that he gave us, but he used spherical plates that we found would be too hard to obtain/machine. So we changed the design considerably. There are two alignment holes in the outer deflection electrode. Machining of the spherical surfaces was done at the Nevis machine shop with a CNC mill. Here's how they looked before gold coating: [[Image(electrodes before coating.jpg, 25%)]]
    5356
    5457== Faraday Cup == #Cup
    55 We have a [http://en.wikipedia.org/wiki/Faraday_cup Faraday Cup] following the design of [http://www.sciencedirect.com/science?_ob=ArticleURL&_udi=B6TJM-4D4CWC0-8&_user=18704&_coverDate=01%2F01%2F2005&_rdoc=1&_fmt=&_orig=search&_sort=d&view=c&_acct=C000002018&_version=1&_urlVersion=0&_userid=18704&md5=743151c7e2ac0d3c3e2105fabf4666b9 Thomas et al.] (Nucl. Instrum. Meth. A '''536''' p.11-21, 2005) that should allow for accurate current readings at low repeller voltages due to nifty design.
     58We have a [http://en.wikipedia.org/wiki/Faraday_cup Faraday Cup] following the design of [http://www.sciencedirect.com/science?_ob=ArticleURL&_udi=B6TJM-4D4CWC0-8&_user=18704&_coverDate=01%2F01%2F2005&_rdoc=1&_fmt=&_orig=search&_sort=d&view=c&_acct=C000002018&_version=1&_urlVersion=0&_userid=18704&md5=743151c7e2ac0d3c3e2105fabf4666b9 Thomas et al.] (Nucl. Instrum. Meth. A '''536''' p.11-21, 2005) that should allow for accurate current readings at low repeller voltages due to nifty design. We have also made a modification to the design to have a shorter, retractable cup with the same entrance opening.
     59[[Image(retractable cup.jpg, 25%)]]