    1010splashed off by H atoms crashing into a collision plate. The plate is within a tube which will have a positive potential
    1111on it (+300-400 V). The secondary electrons are sucked onto this tube and the resulting positive current is read off of
    12 the collsion plate. The postitive tube has to be protected by the incident beam, so a grounded shielding tube protects the
    13 postive tube from the beam.
     12the collsion plate. The postitive electrode tube has to be protected by the incident beam, so a grounded shielding tube protects the
     13postive tube from the beam. The cup must be calibrated because with different vacuum conditions the amount of electrons that come off
     14of the collision plate can change.
     15
     16The length of the positive electrode tube should be at least 2 times the diameter.
     17This cup needs to be calibrated for the number of secndary electrons to incident neutral particles.
     18This ratio is referred to as Gamma. Typical values of Gamma in other designs range from 1 to 1.5.
     19
    1420