    66Here are the drawings for a neutral cup that will be installed soon that
    77will be installed to monitor the neutral H atom beam.
     8
     9The Neutral cup measures our neutral H atom current. This is done by measuring the secondary electrons that are
     10splashed off by H atoms crashing into a collision plate. The plate is within a tube which will have a positive potential
     11on it (+300-400 V). The secondary electrons are sucked onto this tube and the resulting positive current is read off of
     12the collsion plate. The postitive tube has to be protected by the incident beam, so a grounded shielding tube protects the
     13postive tube from the beam.
     14
    815[[br]][[br]][[br]][[br]]
    916[[Image(https://docs.astro.columbia.edu/attachment/wiki/SavinGroup/MergedBeams/EndSection/Neutral%20Cup%20side%20view%20with%20notes.jpg?format=raw, 50%)]]